Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-18
MCB functional Block
MCU function
-
Read the registers of peripheral controller to get the status of sub-systems, such as getting the alarm
status, etc.
-
Write the registers of peripheral controller to control the status of sub-systems, such as PIO
(Programmable Input Output), light indicator, buzzer, etc.
-
Sample watchdog bite signal from supervisor circuit to check peripheral is still working
-
Check all switch signals to respond user’s input
-
Communicate with Control Panel board to get the key input and control the LED output
-
A/D convert all analog signal, such as shutter location, temperature of oil tank, mA feedback, KV
feedback, etc.
-
Communicate with DAC to control mA and KV output
-
Communication with Workstation
-
Communication with peripheral controller to report the working status of MCU
Peripheral controller function
-
Sample the status of input pins, and record them into status registers polled by MCU
-
Set output pins according to the control registers accessed by MCU
-
Generate interrupt signal of alarm to MCU in order to request the processing
-
Generate exposure control signals
-
Check switch stuck error
-
Control mAS control signals
Operational amplifier circuit function
-
Amplify or adjust input analog signals, such as kV feedback, mA feedback, temperature, shutter
location sample signal, and output those signals to the ADC unit (Analog Digital Converter)
-
Amplify or adjust analog signals output from DAC (Digital Analog Converter), such as kV setting, mA
setting, mA preheat, and output those signals to other sub-systems
-
mA PI control
Transceiver
-
Transceiver circuit is responsible for regulating voltage level and supplying driving current.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...