System Overview
2-23
signals are output to KV control board and filament control board through DAC and operational amplifier
circuit or PI circuit sequentially.
Manual kV Setting
Control panel samples the manual kV setting value, and transports it to MCU by SPI bus. MCU looks up the
responding data and sets DAC circuit. Operational amplifier circuit amplifies kV control signal from DAC, and
outputs signal KV_SET_INPUT to kV control board finally.
Manual mA Setting
Control panel samples the manual mA setting value, and transports it to MCU by SPI bus. MCU looks up the
responding data and sends out signals to DAC circuit. The mA control analog signals, KVC and MAC, are
output to filament control board through DAC and PI circuit sequentially.
mAs Control
The setting value of mA is 20mA (or 10mA) in mAS mode. mA sample signal is transformed to pulse by VFC
circuit. The pulse is used as clock signal of down counter in peripheral controller. When mAS mode is set, the
mAS time is set in down counter as the initial value. While exposure starts, down counter works at the same
time. When the value of down counter reduces to 0, exposure stops.
High Level Fluoro
Logically, there is a table that consists of 4 columns: input kV value, input mA value, output kV value and
output mA value. MCU looks up the table by input kV and mA value, and gets output kV and mA value as kV
and mA setting respectively, then outputs those setting.
Auto mA Calibration
Because valid exposure period in each frame is smaller than 40 ms, response time from MAC signal set out
to mA stable should be within a certain time. Meanwhile, the response time is related with the parameters of
tube. The key parameter is the current of tube before exposure. If it is very near the current in exposure
followed, the response time will be short enough to meet the requirement.
Therefore, the rational is that a preheat table is setup to provide the mA_preheat signal in section 6.3.4
Figure 5, which give out a proper current of tube before exposure, and PI control will control mA stable in
several ms’ in exposure.
Unfortunately, the variation exists among the parameters of tubes. Although the variation is small, unique
preheat table should be setup for each tube. In order to get this unique preheat table, auto mA calibration
process shall be as follows.
- The variation is so small that a draft preheat table is generated from mass experiment, which is usable
for calibration
- MCU sends out mA_preheat signal according to the draft preheat table instead of MAC signal to filament
board
- Sample mA feedback (mA_sample signal in section 6.3.4 Figure 5) with ADC unit of MCU
- Then the relationship between MAC and mA of the tube is gotten
- Interpolation to get a unique preheat table of the tube, in which mA value is normalized.
Exposure Control
Peripheral controller controls the exposure command and camera LIH (last image hold) as well as other
peripheral control signals. Refer to PDL document.
Debug
JTAG of Peripheral controller and proprietary COM port of MCU has reserved for debug.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...