System Overview
2-61
X-Ray/key Switch and Emergency Switch
Key switch is used to control the unauthorized usage of X-ray exposure and vertical lift motor.
Signals from Key switch to Mainframe Control Board:
Name
Description
Attribution Scale
Comment
XRAYOFF The termination which
indicates lift column enable
and x-ray disable
Passive
Open/Close XRAYOFF & COMMON is connected means
XRAYOFF switch has pressed, which represents lift
column function has been selected and exposure
enable has not been selected.
ON
The termination which
indicates lift column and x-
ray are both enable
Passive
Open/Close ON & COMMON is connected means ON switch has
pressed, which represent both lift column and
exposure enable function have been selected.
COMMON Common termination
Passive
-
-
X-rays can be enabled via the footswitch or the Hand switch. Signals from footswitch and Hand switch to MCB
are listed in the following charts. Detailed interconnect information, please refer to chapter 2.2.3
Signals from Hand Switch to MCB:
Name
Description
Attribution
Scale
Pin
Comment
HAND_FLU
Hand switch termination Passive
Open/Close J1-B28 HAND_FLU & HAND_COM is
connected means
HAND_FLU button has
pressed
HAND_HLF_SECURITY Hand security line
termination
Passive
Open/Close J1-C27 -
HAND_HLF
High Level Fluoro hand
switch termination
Passive
Open/Close J1-B7
HAND_HLF & HAND_COM is
connected means
HAND_HLF button has been
pressed
HAND
_HLU_SECURITY
High Level Fluoro hand
security line termination
Passive
Open/Close J1-A22 -
HAND_COM
Common termination
Passive
-
J1-C15 -
Signals from Footswitch to MCB:
Name
Description
Attribution
Scale
Pin
Comment
FOOT_FLU
Foot switch
termination
Passive
Open/Close J1-A27 FOOT_FLU & FOOT_COM is connected
means FOOT_FLU button has been
pressed
FOOT_FLU_SECURITY
Foot security
line
termination
Passive
Open/Close J1-C28 -
FOOT_ HLF
High Level
Fluoro foot
switch
termination
Passive
Open/Close J1-A14 FOOT_HLF & FOOT_COM is connected
means FOOT_HLF button has been
pressed
FOOT_ HLF _SECURITY High Level
Fluoro foot
security line
termination
Passive
Open/Close J1-C31 -
FOOT_BUTTON1
Button 1
Passive
Open/Close J1-A20 FOOT_BUTTON1 & FOOT_COM is
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...