Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-6
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0102804
Warning
Tube is
overheating
Tube overheated. Exposure is
disabled. Please wait for the
tube to cool down before
making another exposure.
1. Check connections and
bundle cable.
2. Check OIL_TEMP signal
(temperature sensor inside
monoblock) to MCB.
3. Check the monoblock
0103001
Warning
Iris stuck
Collimator iris stuck. Please
restart the system. If this issue
persists, please call for service.
IRIS stuck
1.Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3.CCB may be abnormal.
4.MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
abnormal.
0103002
Warning
Leaves
open/close
stuck
Collimator leaves open/close
stuck. Please restart the
system. If this issue persists,
please call for service.
Shutter open/close stuck
1. Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3. CCB may be abnormal.
4. MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
abnormal.
0103003
Warning
Leaves rotation
stuck
Collimator leaves rotation
stuck. Please restart the
system. If this issue persists,
please call for service.
Leaves rotation stuck
1. Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3. CCB may be abnormal.
4. MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
abnormal.
0103004
Warning
IRIS width SPI
channel is
abnormal
Collimator iris monitoring
exception. Iris can’t open or
close. Please restart the
system. If this issue persists,
please call for service.
MCU IRIS width A/D failure,
MCB may be abnormal
0103005
Warning
Leaves width
feedback SPI
channel is
abnormal
Collimator leaves open/close
monitoring exception. Leaves
can’t open or close. Please
restart the system. If this issue
persists, please call for service.
MCU Leaves width A/D
failure, MCB may be
abnormal
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...