Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-26
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0306523
Warning
Failed to load
Exposure config
file
Exposure configuration loading
exception. Some configuration
may be missed. Please restart
the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306524
Warning
Get dose
summary
modality type
failed
Get Dose summary modality
setting failed. Please restart
the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306525
Warning
Setup CD/DVD
for importing
image failed
Import images from CD/DVD
exception. Please try again. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306026
Warning
Get string failed Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0306527
Error
ImageDirectory
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0306528
Warning
Stop
application
exception
Close ImageDirectory
exception. Please restart the
system. If this issue persists,
please call for service.
0306529
Error
ImageDirectory
module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304501
Warning
Start exam
exception
Open Exam Management
failure. Please restart the
system. If this issue persists,
please call for service.
0304510
Warning
Init
XmlFileParseUtil
ity for
ConnectivitySet
up.xml failed
Connectivity configuration
loading failure. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...