0408016
Warning
Can’t detect
UPS
System can’t detect UPS.
Please keep AC plugged in to
avoid data loss. If this issue
persists. Please call for service.
Communication between
UPS and WKS is abnormal.
1. PC USB for UPS may be
abnormal
2.connection between PC
and UPS may be abnormal
3. UPS may be abnormal.
4. Reinstall WKS software
0408018
UPS needs to
be replaced for
aging of battery
UPS battery aged. Please call
service to replace the battery.
0408019
Warning
UPS battery
overloads
UPS battery overloaded. Please
restart the system to avoid
data loss. If this issue persists,
please call for service.
1, PC power supply may be
abnormal
0409501
Error
OSGI boot up
failed
Application modules boot up
failure. Please restart the
system. If this issue persists,
please call for service.
0409502
Error
C module boot
up failed
System modules boot up
failure. Please restart the
system. If this issue persists,
please call for service.
0409503
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0404501
Error
TRU module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0404502
Error
TRU module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...