Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-34
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0310501
Warning
Get image and
zoom
information
from IE failed
Workstation communication
exception. Can’t get image and
zoom roam data from the left
monitor. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310502
Warning
Set zoom to IE
failed
Workstation communication
exception. Can’t set zoom
roam data to the left monitor.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310503
Error
ZoomRoam
module register
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310504
Error
ZoomRoam
module
subscribe sync
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310505
Error
ZoomRoam
module
unsubscribe
sync event
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310506
Warning
ZoomRoam
module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310507
Warning
Init zoom
controller
exception
Zoom Roam init exception.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310508
Warning
Parse data
exception
Parse Zoom Roam data
exception. Display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...