Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-14
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
please call for service.
0115004
Warning
Flash write fault
during beam
alignment
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0115005
Warning
Flash erase
fault during
Collimator
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0115006
Warning
Flash write fault
during DAP
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
Flash write fault during DAP
calibration
1.MCB may be abnormal
0117001
Warning
KV is not
accurate
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the kV calibration.
2. Check the bundle cable
3. Check emergency switch
4. Check the MCB.
5. Check the kV control
board.
6. Replace the monoblock.
0117002
Warning
MA is not
accurate
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the mA calibration.
2. Check the bundle cable
3. Check Hand/ foot switch
signal and emergency
switch.
4. Check the MCB.
5. Check the FDB.
7. Replace the monoblock.
0117003
Warning
MA is very low
(<0.05ma)
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the mA calibration.
2. Check the bundle cable
3. Check emergency switch.
4. Check the MCB.
5. Check the FDB.
6. Replace the monoblock.
0117004
Warning
MA is very
high(>23.5ma)
System execution exception.
Please try again. If this issue
persists, please call for service.
Redo mA calibration
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...