Troubleshooting
6-11
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0108801
Warning
During bootup ,
right foot
switch is
pressed
Right foot switch pressed
during boot up. Please release
the right foot switch.
N/A
0109800
Warning
Key switch is
rotated to "X-
ray Off" position
Key Switch in Xray Off.
Exposure is disabled. To enable
exposure, please turn it to On.
N/A
0109801
Warning
Key switch is
rotated to "Off"
position
Keyswitch in Off. Exposure and
lift column motion are
disabled. To enable exposure
or lift column motion, please
turn it to On.
N/A
0110800
Warning
Emergency
switch is
pressed
Emergency switch pressed.
Exposure and lift column
motion are disabled. To enable
exposure or lift column motion,
please release the switch.
N/A
0112500
Warning
DAP event
buffer overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. Don't follow calibration
procedure to perform
calibration. (Reboot system
and Login Serivice Interface,
directly perform collimator
calibration to resolve this
problem.)
2. MCB may be abnormal
0112501
Warning
DAP parameter
buffer overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. Don't follow calibration
procedure to perform
calibration. (Reboot system
and Login Service Interface,
directly perform collimator
calibration to resolve this
problem.)
2. MCB may be abnormal
0112800
Warning
DAP is not
calibrated
without filter
Dose not calibrated for the
tube without filter option.
Please call service to calibrate
dose.
0112801
Warning
DAP is not
calibrated with
filter
Dose not calibrated for the
tube with filter option. Please
call service to calibrate dose.
0113002
Warning
WKS UART
write fault
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
MCB serial port writes to
WKS fault.
1. MCB may be abnormal.
2. Motherboard may be
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...