Brivo OEC 715/785/865 Mobile C-Arm X-ray Product Service Manual
8-10
8.6.
System Function checks
Check the following areas of the C-Arm and workstation for proper operation.
8.6.1.
Perform Power On Checks
Observe safety procedures when operating motorized features.
Perform the following steps with the system powered up and normally booted.
1.
Connect the integrated cable in the C-Arm connector and lock it. Connect the Footswitch cable to the
C-Arm.
2.
Plug the AC power cord into the receptacle.
3.
Press the system on key located on the C-Arm control panel. Verify that the workstation and C-Arm
complete their software boot and initialization with no errors reported on the workstation and C-Arm
displays.
8.6.2.
C-Arm Checks
Mechanical Movement Test:
Refer to section 3.6.2 for C-arm mechanical movement checks.
If wigwag brake can’t prevent Wig Wag motion effectively, or the lock and release
positions can’t fit as the label indicating, perform below procedure to refresh the force of
wigwag brake.
38.5N<F
orce
of
wigwag brake
<
76.9N
Wigwag brake PM:
1.
Pull up the handle far enough (about 3mm) that it clears the fixed shaft.
2.
Turn the handle anticlockwise
3.
Push down the handle.
WARNING
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...