Troubleshooting
6-9
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0105003
Warning
Filament is over
current by SW
monitor
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check DC power supplier
and fuse F5 on the Mother
board (backplane).
2. Check cables and
connections
3. Check the MCB.
4. Check the FDB.
5. Replace the monoblock.
0106001
Warning
6954 chip
verification is
failed: The
written value is
not same with
the read back
value
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. MAX6954 chip of Display
Board may be abnormal
2. Maybe connection issue
in Cable between Display
Board and Mother board
3. Power supply may be
abnormal
4. MCB SPI channel may be
abnormal
0106002
Warning
7301 chip
verification is
failed: The
written value is
not same with
the read back
value
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. 7301 chip of Display
Board may be abnormal.
2. Maybe connection issue
in Cable between DB and
Mothor board
3. Power supply may be
abnormal
4. Maybe MCB SPI channel
may be abnormal.
0106003
Warning
6954 channel
of SPI bus TX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal
3. DB may be abnormal
0106004
Warning
6954 channel
of SPI bus RX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI write to Display
Board (Nixie tube
) time out:
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal
3. DB may be abnormal
0106005
Warning
7301 channel
of SPI bus TX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI write to Display
Board (Nixie tube
) time out:
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...