Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-32
10.
Recover the X-ray Monoblock assembly.
a.
Fasten the screws of collimator cover with the torque to 1.04Nm.
b.
If the Monoblock is not powered for more than 3 months, perform the warm up
procedure firstly per section 4.5.
c.
Monoblock can be diagnosed without disassembly and reassembly, just need connect
monoblock FRU to system for troubleshooting with MONOBLOCK TEST JIG CABLE (5085790)
d.
Perform mA and kV calibration without MONOBLOCK TEST JIG CABLE (5085790) installed,
otherwise it may cause mA not accurate warning.
7.3.8.
Collimator
1.
Position the C-Arm for easy access.
2.
Remove the collimator cover.
3.
Release the 3 screws (Inch screws,
released by T15 screw driver) of
collimator lead ring.
4.
Remove 4 collimator secure screws.
5.
Replace a new collimator, fasten
4 collimator secure screws and
reinstall the collimator lead ring
by fastening 3 screws removed in
step3.
6.
Power on the system.
7.
Select fluoroscopy mode and
make sure to enable Automatic
Technique.
8.
During the exposure, press the
shutter close and open button to
check if the movement of the leaf
is smooth.
9.
Close the shutter all the way to the limit to check if the movement of the leaf is smooth.
10.
Refer to 7.3.2 and chapter 4 to perform calibrations and checks:
11.
Fasten all screws and recover the collimator cover
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...