17-10
Chapter 17 Specification
Technical Information
Arrhythmia Type
The following arrhythmia alarm results can be displayed on this equipment.
CAUTION
The arrhythmia detection result differs depending on the model type and software version of
the bedside monitor and central telemetry receiver connected to the network.
Arrhythmia
Description
Detection Criteria
(The arrhythmia analysis is performed on the bedside
monitor.)
Asystole
ON
3 sec. to 10 sec., 1 sec.
increments
Cardiac arrest is detected for more than preprogrammed
time.
VF
ON
A random, rapid electrical activity of the heart is detected.
VT (Ventricular Tachycardia)
ON
9 or more continuous VPC beats are detected.
*1
Slow VT
ON, OFF
9 or more continuous VPC beats are detected.
*2
Run (Consecutive VPC)
ON, OFF
2 beats to 8 beats, 1 beat
increments
Continuous VPC exceeding the preprogrammed value (2
beats to 8 beats) is detected.
*3
Couplet (Couplet VPC)
ON, OFF
2 continuous VPC beats are detected.
Pause
ON, OFF
1.5 sec. to 5.0 sec., 0.5 sec.
increments
Cardiac arrest exceeding the preprogrammed duration is
detected.
Bigeminy (Ventricular Bigeminy)
ON, OFF
QRS pattern of V-x-V-x-V-x is detected.
*4
Trigeminy
ON, OFF
QRS pattern of x-x-V-x-x-V is detected.
*4
Frequent (Frequent VPC)
ON, OFF
1 bpm to 50 bpm, 1 beat
increments
VPC exceeding the preprogrammed value is detected within
1 minute.
Tachy(Tachycardia)
ON, OFF
The upper HR alarm limit is exceeded.
Brady (Bradycardia)
ON, OFF
The lower HR alarm limit is exceeded.
Ext Tachy (Extreme Tachycardia)
ON, OFF
22 bpm to 300 bpm,
22 bpm to 25 bpm, 1 beat
increments
25 bpm to 300 bpm, 5 beat
increments
The upper alarm limit of extreme tachycardia is exceeded.
Ext Brady (Extreme Bradycardia)
ON, OFF
20 bpm to 295 bpm
20 bpm to 25 bpm, 1 beat
increments
25 bpm to 295 bpm, 5 beat
increments
The lower alarm limit of extreme bradycardia is exceeded.
R on T (R on T VPC)
ON, OFF
200 ms to 600 ms, 8 ms
increments
VPC with the RR interval of same or less than the set
interval (200 ms to 600 ms) is detected.
Multiform (Multiform VPC)
ON, OFF
2 different forms of VPC beats are detected within 4
minutes.
Vent Rhythm (Ventricular Rhythm)
ON, OFF
Continuous VPC beats with HR below the set value for "HR
Lower Limit for Run" (0 bpm to 100 bpm), and same or
above value of the set beats for Run (2 beats to 8 beats) are
detected.
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......