8-4
Chapter 8 Parameter Setup
ECG
The updated time of the reference waveform will be displayed.
NOTE
While learning arrhythmia, or if VPC is present, it will take more than 16 beats to set
the reference waveform.
3
Set the reference point / measurement point.
1
Slide the
for reference point to right and left.
2
Slide the
for measurement point to right and left.
NOTE
Set the reference point in the range of -240 to 0ms in increments of 10ms from the
peak of QRS to the P wave direction.
Set the measurement point in the range of 0 to 560ms in increments of 10ms from the
peak of QRS to the T wave direction.
ST Level Alarm Setup
The ST upper value and lower value compared with the reference waveform will be set.
1
Select the ST Setup screen.
2
Select [ON]/[OFF] for "ST All Alarm" .
3
Select the lead to set the alarm limit.
The selected lead will be displayed large at the right.
For the lead not selected on the screen, switch the page using
/
.
4
Select [ON]/[OFF] of alarm for the selected lead.
This setup is not available for LW bed.
5
Slide
/
on the right of the bar to set the upper/lower limit.
The upper and lower limit can be set in 1mm / 0.1mV increment.
NOTE
Set the upper limit in the range from -19 mm to +20 mm / -1.9 mV to +2.0 mV. If a value
above +20 mm / +2.0 mV is set, the upper alarm will turn OFF.
Set the lower limit in the range from -20 mm to +19 mm/ -2.0 mV to +1.9 mV. If a value
5
3
2
4
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......