9-3
Chapter 9 Data Review
Graphic Trend
memory error can be considered. In such case, trend data cannot be properly stored or
displayed. Refer to the "Troubleshooting" section.
The trend data is periodically stored to the internal memory, but if the power has been turned
OFF for 5 minutes or more, maximum of 5 seconds of data may be lost. In such case, the
lost data will be displayed as not measured data.
REFERENCE
The displayed data is compressed as follows depending on the display interval.
VPC: Maximum value within the display interval
APNEA: Maximum value within the display interval
Other than above: Latest value within the display interval
For example, if the 24-hour trend for the parameter with minimum resolution of 30 seconds
is displayed, one mark will be displayed for the 12-minute (720-second) data.
If the display resolution is higher than the minimum resolution of the data, the same data is
repeated to match the display resolution.
Refer to the following table for resolution. The resolution will differ depending on the
parameter.
Display Resolution
Time Span
Minimum Resolution
Line Display
Mark Display
10 sec. Sample
30 sec. Sample
10 sec. Sample
30 sec. Sample
20 min
10 sec.
30 sec.
10 sec.
30 sec.
1 hours
10 sec.
30 sec.
30 sec.
30 sec.
2 hours
10 sec.
30 sec.
60 sec.
60 sec.
4 hours
20 sec.
60 sec.
120 sec.
120 sec.
8 hours
40 sec.
120 sec.
240 sec.
240 sec.
12 hours
60 sec.
120 sec.
360 sec.
360 sec.
16 hours
80 sec.
240 sec.
480 sec.
480 sec.
24 hours
120 sec.
240 sec.
720 sec.
720 sec.
Data Resolution
Minimum
Resolution
Parameter
10 sec.
HR, ST, SpO
2
, SpO
2
-2, PR_SpO
2
, PR_SpO
2
-2, BP1, BP2
30 sec.
Other than above (Excluding NIBP
*
)
* Actual measured data will be displayed for NIBP.
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......