8-13
Chapter 8 Parameter Setup
NIBP
message will be displayed, and an alarm sound will be generated.
NIBP
This section explains the procedure for NIBP measurement condition setup.
For procedure to set the NIBP alarm, refer to "Chapter 7 Alarm Function".
NIBP Periodic Measurement
NIBP measurement can be performed automatically at selected time interval or at selected time.
This function is available only for the DS-LANIII bed and the TCON bed.
1
Press the key for "NIBP Auto Mode". ( [Interval]/ [Timer]/ [OFF])
2
When [Interval] is selected, select the interval from [2min.] to [120min.].
The measurement time will be integral multiple of the selected interval time
starting from 0 minute.
Ex.) If the current time is 13:14, the measurement time will be as follows for each
interval time.
2 min: 13:16, 13:18, 13:20, ...
2.5 min: 13:15, 13:17:30, 13:20, ...
3 min: 13:15, 13:18, 13:21, ...
3
When [Timer] is selected, select the measurement starting time. (More than one
selection is possible.)
Auto Mode
NIBP measurement interval (only for DS-LAN bed, TCON bed)
NIBP Alarm
ON/OFF of NIBP alarm, alarm limit for systolic (S), diastolic (D), mean (M) blood pressure.
NIBP Detail Setup
Patient Classification, PR Display, MAP, Time Display
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......