8-12
Chapter 8 Parameter Setup
RESP
Common Setup / Impedance Setup
RR can be measured from the following 3 sources.
Impedance RR
RR measured on the multigas unit
RR measured on the ventilator
There are setup items common to all sources and setup items specific to each source.
Common Setup
1
RR Synchronized Mark
[ON]: The RR synchronized mark will be displayed.
[OFF]: The RR synchronized mark will not be displayed.
2
RR/APNEA Alarm Source
RR/APNEA Alarm Source can be set only on the bedside monitor. The setting will be displayed on the central
monitor.
[Impedance]: RR alarm will be generated based on the impedance respiration curve. Also, the RR
synchronized mark based on impedance respiration will be displayed.
[CO
2
/GAS]: RR alarm will be generated based on the RR measured by the CO
2
module or multigas unit.
Also, the RR synchronized mark based on CO
2
waveform will be displayed for only the wired network beds.
[Vent.]: RR alarm will be generated based on the RR measured by the ventilator. Also, the RR synchronized
mark based on ventilator measurement will be displayed for only the wired network beds.
Impedance Setup
1
CVA Detect
[ON]: When CVA is detected, alarm will generate and message will be displayed.
[OFF]: CVA detection will not be performed.
When the amplitude of the respiration waveform decreases due to causes such as respiratory pause, the ECG
waveform may be superimposed on to the respiration waveform, making the RR equal to the HR. This
condition is called CVA (Cardio-Vascular Artifact), and is detected using the CVA detection function
This function will be effective only when [Impedance] is set as the "RR/APNEA Alarm Source".
If the ECG waveform is superimposed on to the respiration waveform with HR (RR) of 30Bpm or above for
more than 20 seconds (10 seconds for neonates) and if the "CVA Detect" is set to [ON], the <CVA detected>
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......