8-14
Chapter 8 Parameter Setup
NIBP
Detail Setup
The following items can be set on the "NIBP Detail Setup" screen.
1
Patient Classification
The patient classification setting is linked with that on the "Admit/Discharge" screen. The inflation value and
measurement duration will differ according to the patient classification setting.
(
"Entering the Patient Information" P6-2)
WARNING
The patient classification selection influences the precision of the QRS detection and
NIBP measurement. Make sure the correct selection is made.
To perform correct NIBP measurement, appropriate NIBP air hose corresponded to
the set patient classification must be used. (However, if the patient classification is
child, NIBP air hose for adult can be used.)
2
PR Display
[ON]: PR will be displayed.
NOTE
When the DS-8500 system bedside monitor is used, PR will be displayed only. It will
not generate alarm, or be displayed for the tabular trend function.
3
Mean BP (MAP) Display
ON/OFF of the MAP display can be set only on the bedside monitor. The setting will be displayed on the central
monitor.
CAUTION
If the mean BP (MAP) value is not displayed, the mean BP (MAP) alarm will not be
First Page
Patient Classification, PR Display
Second Page
MAP
Third Page
Time Display
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......