7-24
Chapter 7 Alarm Function
Alarm History
Stored Events for Alarm History
The following events will be stored for alarm history.
Classification
Item
Stored Event, Data
Each Bed
Numeric Data Alarm/Arrhythmia
Alarm
Time, duration of alarm occurrence
Numeric data, alarm setting at alarm occurrence
Alarm Level
Equipment Alarm
Time, duration of alarm occurrence
Alarm Level
Low Battery Symbol
Time, duration of low battery symbol display
Telemetry Reception
Time, duration of "Too Far Alarm" occurrence
Change of Alarm Settings
Change of alarm ON/OFF, upper/lower limit ON/OFF, upper/lower
limit value
Control Source (Manual*, DS-LAN, TCON)
Alarm Silence
Starting of Alarm Silence, Control Source (Manual*, DS-LAN)
Alarm Suspend
Starting of Alarm Suspend, Control Source (Manual*, DS-LAN)
Canceling of Alarm Suspend, Control Source (Manual*, DS-LAN)
Alarm Sound Suspend
Time of alarm sound suspend start/cancel
Discharge, Bed Transfer/
Exchange
Control Source (Manual*, DS-LAN, EMR, etc.)
Monitor Suspend
Monitor Suspend, Monitor Resume
All Beds
Alarm Sound Level
Change of alarm sound level
Change of Alarm Level
Change of alarm level setting
Change of Alarm Silence Time
Change of alarm silence time setting
Change of Alarm Suspend Time
Change of alarm suspend time setting
Change of Alarm Sound Suspend
Setting
Change of alarm sound suspend setting
*: "Manual" for the control source indicates that the operation is performed on this equipment.
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......