xxix
Compliance to the Electromagnetic Immunity (1)
The DS-8400 System is intended for use in the electromagnetic environment specified below. The customer or the
user of the DS-8400 System should assure that it is used in such an environment.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6kV: contact
±8kV: air
±6kV: contact
±8kV: air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2kV: power supply lines
±1kV: input/output lines
±2kV: power supply lines
±1kV: input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV: differential mode
±2kV:common mode
±1kV: differential mode
±2kV:common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
<5% U
T
(>95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Mains power quality should be that of
a typical commercial or hospital
environment.
If it is required to continuously
operate the DS-8400 System during
power failure, it is recommended to
operate on an uninterrupted power
supply.
Power Frequency (50/60Hz)
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
*: U
T
is the AC mains voltage prior to application of the test level.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...