![Fukuda Denshi DS-8400 Скачать руководство пользователя страница 184](http://html.mh-extra.com/html/fukuda-denshi/ds-8400/ds-8400_service-manual_2344852184.webp)
9-16
Non-Invasive Blood Pressure
The cuff is not inflated although the pump is operating.
Cause 1
The air hose is not firmly connected, and the air is leaking.
Solution
Check if the air hose is properly connected.
Cause 2
The cuff size does not match the selected patient type.
Solution
Use the cuff with correct size for the selected patient type.
The pump is not operating.
Cause
The air hose is disconnected from the NIBP connector.
Solution
Check if the air hose is properly connected.
The measurement data is displayed as "---" .
Cause 1
The measurement accuracy is not reliable due to body motion artifact.
Solution
During the measurement, have the patient stay still.
Cause 2
The pulse is too small to acquire reliable measurement accuracy.
Solution
Check if the cuff application is proper, and if the cuff size corresponds with the selected patient type.
Cause 3
The air hose is disconnected.
Solution
Check if the air hose is tightly connected, and then measure again. If the same message is displayed again, air
leakage inside the HS-8000/DS-8007 can be considered.
<Check NIBP cuff, hose> is displayed.
Cause 1
The connection between the cuff and air hose or the air hose and NIBP connector is loose or disconnected.
Solution
If the connection is loose or disconnected, securely connect it and perform the measurement again.
If the same message is displayed again, internal air leakage can be considered.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...