9-13
Cause 4
The sensor is used beyond its expected life.
Solution
Replace the sensor.
The X-Cal function automatically monitors the Active Monitoring Time (actual time of SpO
2
monitoring) for each sensor and cable.
Even if the sensor is used beyond its expected life, the measurement will not cease unless
the power is turned OFF, sensor is disconnected from the cable, cable is disconnected from
the monitor, or the sensor is reattached.
When a measurement with a sensor that has reached its end of life is suspended for certain
amount of time, and resumed with the same sensor, a message to replace the sensor will
be displayed.
Depending on the equipment, some sensors may not be recognized.
<SpO
2
Check Sensor Attach.> is displayed.
Cause 1
The sensor is detached from the patient.
Solution 1
Check if the sensor is properly attached to the patient.
Solution 2
Check that the light emitting and receiving parts of the sensor LED are aligned.
Cause 2
The sensor is exposed to too much ambient light. The detecting part of the sensor is not covered appropriately.
Solution 1
Turn down or turn off the light.
Solution 2
Avoid the sensor from exposure to ambient light.
Solution 3
Relocate the sensor position.
<SpO
2
Low Perfusion> is displayed.
Cause
The amplitude of the pulse waveform is too low, or the sensor is not positioned correctly.
Solution
Check that the light emitting and receiving parts of the sensor LED are aligned.
<Low Confidence> is displayed.
Cause
The amplitude of the pulse waveform is too low, or the sensor is not positioned correctly.
Solution
Check that the light emitting and receiving parts of the sensor LED are aligned.
NOTE
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...