xxi
not be displayed.
If the HR/PR source is set to [SpO
2
], and if SpO
2
waveform/numeric data is set to [Disp. OFF], the PR value
will not be displayed.
If the RR source is set to [CO
2
/GAS], and if CO
2
waveform/numeric data is set to [Disp. OFF], the RR value
will not be displayed.
If the RR source is set to [CO
2
/GAS], and if GAS waveform/numeric data is set to [Disp. OFF], the RR value
will not be displayed.
Do not set the same remote control ID to multiple monitors in the same floor. Otherwise, the remote control
operation may control multiple monitors at the same time.
After the remote control setup, check that the remote control unit is properly operating.
If the time/date is not correctly set, or changed during monitoring, malfunction may occur with NIBP
measurement, periodic printing, trend, NIBP list data, and age calculation from the birth date.
If the time/date is changed, the time/date for all the saved patient data (trend, list, recall, etc.) will also change.
The printed time/date before changing and the displayed time/date after changing will differ.Also, the data
transmitted to the central monitor before the time/date is changed will be displayed on the central monitor with
the previous time/date.
Precautions about the Multigas Unit Data Monitoring (MGU-800/MGU-810 Series)
When the multigas unit (MGU-800/MGU-810 series) and HPD-800/HPD-810/HPD-820, HCP-800/HCP-810/
HCP-820 are simultaneously used, the CO
2
concentration measurement will be performed by the equipment
selected for the "CO
2
Source Priority" under ( [Menu] > "Parameter" [CO
2
]) .
The MGU-800/MGU-810 series require warm up of about 10 minutes to correctly measure the data.
If the power supply is interrupted due to power failure, etc., MGU-800/810 series multigas unit will initialize
and enter into warm-up mode even if the power interruption is within 30 seconds.
About the Gas Calibration
The gas calibration will automatically start when the MGU-800/810 series multigas unit is connected.
After the warm-up completes, gas calibration will be performed every 4 hours during stable operation.
During warm-up, gas calibration interval will become shorter than during normal operation.
During gas calibration, measurement data will not be updated.
Calibration gas is not required during gas calibration.
Dispose of calibration gas according to the regulation of each medical institution.
Make sure the sampling line and flow sensor is securely connected to prevent any leakage.
The gas leakage inside or outside the equipment will cause measurement error. Make sure to check for leakage
before usage.
An environment with alcoholic vapor may adversely affect the measurement readings.
CO
2
, N
2
O or anesthetic agent in the atmosphere around the MGU-800/810 series may adversely affect the
measurement readings.
SPIRO and ventilator cannot be used simultaneously.
During the warming up process, the date of the last measurement accuracy check cannot be updated. Perform
the measurement accuracy check after the warming up process is completed.
If the accuracy check is performed using a low pressure gas, the accuracy of gas measurement will be reduced.
Make sure to perform the accuracy check using the specified calibration gas before its expiration date.
If the error persists, refer to Fukuda Denshi.
About the MGU-810 Series
The adult flow sensor dead space is 6.9 mL and the flow resistance is 1.8 cmH
2
O at 60 L/min.
The pediatric flow sensor dead space is 0.75 mL and the flow resistance is 0.9 cmH
2
O at 10 L/min.
Adjust ventilation accordingly.
To prevent condensation, the patient breathing circuit, flow sensor and pressure tubing should not be directly
exposed to cooling equipment such as fans or cooling blankets.
Leakage of gas from the patient breathing system may occur if the pressure or gas sampling lines are not
connected to the MGU-810.
CAUTION
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...