10-6
Maintenance (DS-LAN) :
Since these displays are used for development, do not press these keys. If
accidentally displayed, immediately return to the previous display.
C A U T I O N
Never operate the display for development. It may damage the equipment.
EEPROM (HS-8000) :
EEPROM (Ext.)
:
These are status displays for development.
LOG 1 (HS-8000)
: Records the time and information if abnormal restart and exception handling
occur in the HS-8000, and stores the information when equipment failure
occurs.
Analog Parameter
: Measured value of a parameter under measuring etc. appears.
RTC Setup
: This is the factory initial setting. During normal operation, do not press this key.
Set the time to the reference clock of the main unit.
The normal time setting should always be performed on the "Time/Date"
setup under the "Settings" menu.
All of integrated operating time and recording time of trend and recall will be
changed when you change this time, unlike the clock setup on machine setup.
Hence, you need to pay attentions since the initialization of the system,
EEPROM, etc. will be required.
Set up the RTC time only when the Lithium Battery, RTC, or CPU board is
replaced.
Usually, this setup is not changed after setup at factory.
Make sure to initialize the system and operation time of EEPROM after setup.
WDT TEST
: WDT test used for factory inspection. During normal operation, do not press
this key. All of Patient data, etc. that have been previously set will be cleared.
Demo Start
: A demonstration is carried out.
This Demo Start key appears when the rotary switch on the side of the main
body is set to “2” or “7”.
After the demonstration, turn off the power and return the rotary switch to “0”.
All of Patient data, etc. that have been set until now will be cleared.
If the ECG cable is inserted, the key cannot be pressed.
Analogue Adjust
: This adjusts the analog board.
A special tool is required to adjust the analog board.
For details, refer to the section of analog adjustment screen.
Initialization
: All of Patient data, etc. that have been set until now will be cleared.
Alarm Log
: The log of the generated alarm will be displayed.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...