9-44
<GAS Mixed Agents Detection> is displayed.
Cause
More than one halogenated anesthetic gas exists.
Solution 1
Make sure that multiple anesthetic gases are not used.
Make sure that the anesthetic gas carburetor setting is correct.
<GAS Zeroing Failed> is displayed.
Cause
The zero calibration process has not been properly completed.
Solution
Perform the manual zero calibration again.
<SPIRO Unit Failure> is displayed.
Cause
The hardware failure of the SPIRO unit was detected.
Solution
Contact Fukuda Denshi.
<SPIRO Check FlowSensor Class> is displayed.
Cause 1
The flow sensor is disconnected or not securely connected.
Solution
Make sure that the flow sensor is securely connected.
Cause 2
The flow sensor is damaged.
Solution
Replace the flow sensor.
Cause 3
The used flow sensor does not correspond to the patient classification setting on the monitor.
Solution
Make sure that the used flow sensor corresponds to the patient classification setting.
When the patient classification is adult, use the flow sensor intended for adult.
When the patient classification is "Child" or "Neonate", use the flow sensor intended for pediatric.
<SPIRO Check Flow Sensor> is displayed.
Cause
If the flow sensor is disconnected during multigas monitoring, the message will be displayed.
Solution 1
To cease multigas monitoring, press the [Alarm Silence] key. The message will disappear, and the alarm will be
silenced.
Solution 2
To continue monitoring, plug in the flow sensor. The message will disappear, and the alarm will be silenced.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...