![Fukuda Denshi DS-8400 Скачать руководство пользователя страница 266](http://html.mh-extra.com/html/fukuda-denshi/ds-8400/ds-8400_service-manual_2344852266.webp)
11-8
No Check
Items
Check
Procedure
Criteria
15. Periodic Replacement Parts, Aftertreatment
01 NIBP
Unit
Check the number of measurements or date of
first usage.
It should not exceed 100,000 times, or six
(6) years from the first usage.
02
Lithium-Ion Battery
Pack BTO-005
Check the number of charging/discharging
times or the first usage date.
The number of charging/discharging
times should not exceed 300 times, or
usage duration should not exceed one
year.
03
CO
2
Unit
(When HCP-810/
HCP-820 is used)
Check the usage hours.
It should not exceed 30,000 hours.
04
Restore Backup Data
After the check procedure, restore the setup
data from the SD card or USB memory.
If the setting information before the check
procedure have been written down, restore the
settings to original state as written.
The settings should be properly
restored to original state.
No Check
Items
Check
Procedure
Criteria
16. Electrical safety
01 Earth
Leakage
Current (NC)
Measure the earth leakage current under
normal condition using a leak measurement
safety tester.
Test according to the test method of IEC 60601-1.
Earth Leakage Current (NC)
Should be 5mA or less
02 Earth
Leakage
Current (SFC)
Measure the earth leakage current under single
failure condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
Earth Leakage Current (SFC)
Should be 10mA or less
03
Touch Current (NC)
Measure the touch current under normal
condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
Touch Current (NC)
Should be 100
μ
A or less
04
Touch Current (SFC)
Measure the touch current under single failure
condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
Touch Current (NC)
Should be 500
μ
A or less
05 Patient
Leakage
Current (NC)
Measure the patient leakage current under
normal condition using a leak measurement
safety tester.
Test according to the test method of IEC 60601-1.
[AC/DC] Type CF Applied Part
Patient Leakage Current (NC)
Should be10
μ
A or less
06 Patient
Leakage
Current (SFC)
Measure the patient leakage current under
single failure condition using a leak
measurement safety tester.
Test according to the test method of IEC 60601-1.
[AC/DC] Type CF Applied Part
Patient Leakage Current (SFC)
Should be50
μ
A or less
07 Total
Patient
Leakage
Current (NC)
Measure the total patient leakage current under
normal condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
[AC/DC] Type CF Applied Part
Patient Leakage Current (NC)
Should be 50
μ
A or less
08 Total
Patient
Leakage
Current (SFC)
Measure the total patient leakage current under
single failure condition using a leak measurement
safety tester.
Test according to the test method of IEC 60601-1.
[AC/DC] Type CF Applied Part
Patient Leakage Current (SFC)
Should be 100
μ
A or less
09 Patient
Auxiliary
Current (NC)
Measure the patient auxiliary current (NC)
under normal condition using a leak
measurement safety tester.
Test according to the test method of IEC 60601-1.
Patient Auxiliary Current (NC)
Should be10
μ
A or less
10 Patient
Auxiliary
Current (SFC)
Measure the patient auxiliary current (SFC)
under single failure condition using a leak
measurement safety tester.
Test according to the test method of IEC 60601-1.
Patient Auxiliary Current (SFC)
Should be 50
μ
A or less
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...