9-49
<Artifacts> is displayed.
Situation: The signal quality is less than half of the level desirable for optimal monitoring conditions.
This message may occur as the results of artifact (non-EEG signal) such as those generated
from motion (patient movement or eye blinks) or the presence of electrocautery, warming
blankets, or other devices.
Cause 1
Artifact, such as those generated by motion or eyeblinks, is causing loss of EEG recognition.
Solution
If <Artifacts> appears on the display, attempt to identify and eliminate artifact source.
Cause 2
EMG bar indicates electrical activity that may be interfering with EEG recognition.
Solution
If EMG bar is illuminated, attempt to determine and eliminate cause.
Cause 3
The patient interface cable (PIC cable) is defective.
Solution
Make sure that sensor check passes. If not, replace the patient interface cable (PIC cable).
Cause 4
The BISx is defective.
Solution
Replace the BISx.
<BIS SQI < 15%> is displayed.
Situation: The signal quality is too low to accurately calculate a BIS value.
The BIS value and other trend variables that are adversely affected by artifact are not displayed.
This message may occur as the results of artifact (non-EEG signal) such as those generated
from motion (patient movement or eye blinks) or the presence of electrocautery, warming
blankets, or other devices.
Cause 1
Artifact, such as those generated by motion or eyeblinks, is causing loss of EEG recognition.
Solution
If <BIS SQI < 15%> appears on the display, attempt to identify and eliminate artifact source.
Cause 2
EMG bar indicates electrical activity that may be interfering with EEG recognition.
Solution
If EMG bar is illuminated, attempt to determine and eliminate cause.
Cause 3
The patient interface cable (PIC cable) is defective.
Solution
Make sure that sensor check passes. If not, replace the patient interface cable (PIC cable).
NOTE
NOTE
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...