![Fukuda Denshi DS-8400 Скачать руководство пользователя страница 186](http://html.mh-extra.com/html/fukuda-denshi/ds-8400/ds-8400_service-manual_2344852186.webp)
9-18
C02-02 The air hose was disconnected from the NIBP connector during the measurement.
Cause
The air hose was disconnected from the NIBP connector during the measurement.
Solution
Connect the air hose to the NIBP connector, and then measure again.
C03-xx The exhaust ventilation has ceased, or the target deflation speed was not achieved.
Cause 1
During measurement, an artifact such as body motion may have interfered.
Solution
Keep the patient still as much as possible, and measure while the patient is not moving. When performing the
measurement during surgery, avoid artifact caused by the surgery.
Cause 2
During the measurement, air hose was bent or occluded by the compression.
Solution
Make sure that the air hose is not bent or compressed before the measurement.
If the error persists and C03-xx error is frequently displayed, contact Fukuda Denshi and notify the error code.
C04-xx The cuff inflation was insufficient for the patient's blood pressure.
Cause
The blood pressure has significantly increased from the previous measurement.
Solution
Check the cuff application and size and perform the manual measurement.
C06-xx The pulse signal detected during the measurement was unstable.
Cause 1
During the measurement, the patient has trembled or moved.
Solution
Keep the patient still as much as possible, and measure while the patient is not trembling or moving.
Cause 2
Arrhythmia has frequently occurred during the measurement.
Solution
If arrhythmia occurs many times, correct measurement cannot be performed. Measure when arrhythmia is not
frequently occurring.
C07-00 The measurement time has exceeded the allowable time.
Cause
Measurement is automatically repeated due to body motion or insufficient inflation.
Solution
Check the cuff application and size, and measure while keeping the patient still as much as possible.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...