![Fukuda Denshi DS-8400 Скачать руководство пользователя страница 264](http://html.mh-extra.com/html/fukuda-denshi/ds-8400/ds-8400_service-manual_2344852264.webp)
11-6
No Check
Items
Check
Procedure
Criteria
04 Alarm
Sound/
Operating Sound
On the "Tone/Volume" menu, check the alarm
sound.
Alarm sound should generate with proper
volume.
There should be no beat noise.
05
Date/Time
Check the year, month, day, and time on the
display.
The year, month, day, and time should be
correctly displayed.
06 Printing
Status
(When HR-800 is
used)
Perform test printing on the maintenance menu.
Visually check the printing condition and also if
there are thin or missing points.
The printed characters should be clear and
legible.
07 Paper
Speed
(When HR-800 is
used)
Perform test printing on the maintenance menu.
Check by measuring the length of printed grid.
Error should be within ±3% for 25 mm/sec
and 50 mm/sec waveform traces.
08 Telemetry
Transmission
Perform telemetry transmission, and check the
reception condition and waveform on the
receiver side.
Correct waveforms and numeric data should
be displayed and receiving condition should
be stable.
No Check
Items
Check
Procedure
Criteria
5. ECG
01
ECG Display Size
Input square wave of 1 mV amplitude from the
simulator, and check the displayed waveform on
the monitor with a waveform size of x1.
The amplitude of the displayed waveform
should be within 10 mm ± 1 mm.
02
Heart Rate Display
Accuracy
Set the heart rate to 60 bpm on the simulator,
and check the displayed heart rate on the
monitor.
The displayed heart rate should be within 60
bpm ± 3 bpm.
03 Lead
OFF
Remove each electrode from the simulator and
check that "Lead OFF" message is displayed.
The "Lead OFF" message for the removed
electrode should be displayed.
No Check
Items
Check
Procedure
Criteria
6. Respiration (Impedance Measurement)
01 Respiration
Rate
display accuracy
Input the respiration signal of 20 Bpm with the
following setting from the simulator, and check
the displayed RR value on the monitor.
Baseline Impedance : 1500
Ω
,
Detection Lead: II (F/LL), Respiration
Waveform: Normal, Amplitude: 1
Ω
The displayed RR should be within 20 Bpm ±
3 Bpm.
No Check
Items
Check
Procedure
Criteria
7. Invasive Blood Pressure
01
BP Zero
Input the blood pressure of 0 mmHg from the
simulator, perform zero balance, and check the
BP value on the monitor.
Zero balance should be properly performed,
and the displayed BP value should be within
0 mmHg ± 1 mmHg.
02
BP value display
accuracy
Set the BP value to 250 mmHg on the simulator,
and check the displayed BP value on the
monitor.
The displayed BP value should be within 250
mmHg ± 5 mmHg.
No Check
Items
Check
Procedure
Criteria
8. SpO
2
01 Oxygen
Saturation
Display Accuracy
Input the oxygen saturation signal of 90%SpO
2
from the simulator, and check the displayed
oxygen saturation value on the monitor.
The displayed oxygen saturation value
should be within 90%SpO
2
± 2%.
02
Pulse Rate Display
Accuracy
Input the pulse rate signal of 60 bpm from the
simulator, and check the displayed HR value on
the monitor.
The displayed heart rate should be within 60
bpm ± 3 bpm.
Содержание DS-8400
Страница 4: ......
Страница 36: ...xxxii ...
Страница 46: ...1 6 ...
Страница 74: ...3 4 ...
Страница 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Страница 120: ...5 4 ...
Страница 128: ...6 8 ...
Страница 130: ...7 2 DSC 8410 CPU Unit BLK ...
Страница 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Страница 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Страница 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Страница 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Страница 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Страница 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Страница 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Страница 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Страница 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Страница 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Страница 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Страница 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Страница 168: ...8 6 ...
Страница 222: ...9 54 ...
Страница 258: ...10 36 ...
Страница 272: ...11 14 ...
Страница 274: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101420 201706 ...