Chapter 15: Appendix
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
237
15.6 Collecting a sample for microbiological analysis
The sampling port is the dialysis water coupling.
15.6.1 Preparation
Have a cooled shipping box available.
The reverse osmosis system must be operated in the
RINSE
or
SUPPLY
mode for at least 20 minutes before the sample is
collected.
While collecting the sample, the reverse osmosis system must be in
the
RINSE
or
SUPPLY
program.
Disconnect the dialysis water connection from the dialysis device,
and perform the procedure described for collecting a sample at the
dialysis water coupling.
15.6.2 Accessories, equipment
The following equipment is recommended by the manufacturer:
– Rubber gloves
– Alcohol-based hand disinfectant
For the chemical sampling, use the sample containers provided by the
laboratory. The bag with adapter (part number: G03000832) can be
used as dialysis water sampling equipment.
Dialysis water coupling
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...