![Fresenius Medical Care AquaA Скачать руководство пользователя страница 205](http://html.mh-extra.com/html/fresenius-medical-care/aquaa/aquaa_instructions-for-use-manual_2331674205.webp)
Chapter 14: Options
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
205
Cooling ring main
Module heat disinfection
Example temperature graph during module heat disinfection:
The temperature graph shows the temperature over time during module
heat disinfection:
If an
AquaA2
is connected, the membranes of the
AquaA2
will be
included in the heat disinfection.
During the first part of this phase, the ring
main is cooled to a temperature below 35 °C
by discarding hot product water.
During the second part of this phase, rinsing
of the ring main is started. Here, dialysis
water is discarded on the
RingBase
for a
preprogrammed duration. This part of the
“Cooling ring main” phase can be stopped
prematurely at any time using the
Cancel
button.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...