12.4 Electrical supply
An overvoltage protection device must be installed to prevent damage
to the fuse in the power distribution box of the
AquaA
system. This can
occur when a surge impulse is caused by an atmospheric source, such
as a thunderstorm, or by an unstable power supply.
When using fuses, replacement is recommended every 24 months as
part of the periodic maintenance (MA) procedures.
The use of 3-pole circuit breakers is recommended.
12.5 Fuses
The following is a list of the fuses installed in the
AquaA
:
Warning
Risk of injury caused by electric shock
Without a protective ground connection, there is a risk of electric shock.
Always connect the device to a power supply network with a
protective earth conductor.
System type
900H/1000/1800H/2000
2700H/3000/3600H/4000
Line voltage
208 V, 60 Hz
Power supply
hardwired
Protection
20 A (for
AquaA
900H/1000/1800H/2000)
32 A (for
AquaA
2700H/3000/3600H/4000)
Tripping characteristic C, D, K or comparable
Power consumption
6.0 kVA
9.6 kVA
Power line impedance
< (0.24 + j0.15) ohm
Note
– A residual current device (RCD) or other suitable measure must be
provided so that the conditions for preventing interruptions of the
neutral conductor are fulfilled.
– The manufacturer recommends using a residual current device
(RCD) which operates at 30 mA.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...