Chapter 12: Specifications
166
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
6 V
rms
in ISM bands
between 150 kHz
and 80 MHz
3 V
rms
150 kHz to 80 MHz
6 V
rms
in ISM bands
between 150 kHz
and 80 MHz
Portable and mobile radio-frequency
telecommunication devices (radio devices
including their accessories, such as
antenna cables and external antennas)
should not be used at a distance of less
than 30 cm (12 inches) to the
AquaA
.
Non-compliance may result in impairment
in the performance of the device.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
9 V/m
704 to 787 MHz
5100 to 5800 MHz
27 V/m
380 to 390 MHz
28 V/m
430 to 470 MHz
800 to 960 MHz
1700 to 1990 MHz
2400 to 2570 MHz
3 V/m
80 MHz to 2.7 GHz
9 V/m
704 to 787 MHz
5100 to 5800 MHz
27 V/m
380 to 390 MHz
28 V/m
430 to 470 MHz
800 to 960 MHz
1700 to 1990 MHz
2400 to 2570 MHz
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
AquaA
device is used
exceeds the applicable RF compliance level above, the
AquaA
device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the
AquaA
device.
Guidance and manufacturer's declaration – electromagnetic immunity
The
AquaA
device is intended for use in the electromagnetic environment specified below. The customer or the
user of the
AquaA
device should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test
level
Compliance level
Electromagnetic environment –
guidance
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...