Chapter 12: Specifications
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
177
12.12 Specifications–AquaHT
Dimensions and weight
Dimensions
Height
1840 mm
Width
610 mm (at the tank 800 mm)
Depth
1200 mm (incl. piping 1410)
Weight
empty
200 kg
filled
620 kg
Device data
Heater output
Max. 19.5 kW
Tank volume
adjustable between 100 and 380 liters
Tank temperature
adjustable from 65 to 85 °C
Ring main temperature
(heat disinfection)
adjustable from 60 to 87 °C
Membrane temperature
(heat disinfection)
adjustable from 60 to 82 °C
Maximum pressure
max. 6 bar
Maximum length of ring
main
Heat disinfection cycles
for ring main heat disinfection:
– Unlimited
for module heat disinfection:
– 160 cycles
Note
The length of the ring mains is restricted as follows:
– Ring length per ring main
max. 250 m
– With three ring mains
max. 600 m
in total
– The ring mains must have thermal insulation.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...