Chapter 12: Specifications
170
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
Service/diagnostics
For inhouse computer diagnosis.
Port:
RJ45
Alarm output
For the connection of an external indicator (staff call or remote control).
(potential-free alarm output, alternating contact maximum 24 V/24 W).
Warning output
For the connection of an external indicator (staff call or remote control).
(potential-free alternating contact maximum 24 V/24 W).
Emergency operation
output
For the connection of an external indicator (staff call). (potential-free
alternating contact maximum 24 V/24 W).
Standby output
For the connection of an external indicator (staff call or remote control).
(potential-free alternating contact maximum 24 V/24 W).
Supply output
For the connection of an external indicator (staff call or remote control).
(potential-free alternating contact maximum 24 V/24 W).
Rinse output
For the connection of an external indicator (staff call). (potential-free
alternating contact maximum 24 V/24 W).
Disinfection output
For the connection of an external indicator (staff call). (potential-free
alternating contact maximum 24 V/24 W).
Heat disinfection output
For the connection of an external indicator (staff call). (potential-free
alternating contact maximum 24 V/24 W).
System inputs
Standby input
Control input: external switchgear must have an electric strength of at
least 4 kV.
Supply input
Control input: external switchgear must have an electric strength of at
least 4 kV.
Warning
Risk for the patient as a result of corrupted data
Data corruption or data loss caused by the network and the server
software cannot be detected by the device. This can lead to
malfunctions.
The system installer must ensure that device data is processed
securely, e.g., in PC software applications.
The network operator must ensure that any data transferred without
encryption is protected.
Tip
The system inputs are to be used for the remote control function.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...