Chapter 5: Alarms
114
Fresenius Medical Care
AquaA
IFU-EN 06A-2021
5.5 Error category 02–violation of alarm limits
An error code marked with „*“ denotes a self-confirming message.
Error code
Visual
indica-
tor
Messages
Cause
Action required
F–02–01–01
Signal:
red
FAILURE:
Permeate cond.
alarm limit
exceeded
– The dialysis water
conductivity has exceeded
the specified alarm limit.
– Conductivity sensor CD-P
defective
– Measuring transducer B4
defective
This failure is confirmed
automatically when the
value drops below the
alarm limit. The
message will continue to
be shown on the
display, however.
Contact the service
department
F–02–01–02
Signal:
red
FAILURE:
Permeate temp.
alarm limit
exceeded
– The dialysis water
temperature has exceeded
the specified alarm limit.
– Conductivity sensor CD-P
defective
– Measuring transducer B4
defective
This failure is confirmed
automatically when the
value drops below the
alarm limit. The
message will continue to
be shown on the
display, however.
Contact the service
department
F–02–01–03
Signal:
red
FAILURE:
Permeate press.
alarm limit
exceeded
– The dialysis water pressure
has exceeded the specified
alarm limit.
– Pressure sensor P-P
defective
– Measuring transducer B4
defective
Contact the service
department
F–02–01–04
Signal:
red
FAILURE:
Conc. pressure
alarm limit
exceeded
– The concentrate pressure
has exceeded the specified
alarm limit.
– Pressure sensor P-C
defective
– Measuring transducer B4
defective
Contact the service
department
F–02–01–05
Signal:
red
FAILURE:
run-dry protection,
pump stop
– The level in the break tank
has dropped to
NIV1
.
– Water inlet pressure or flow
too low
Check water supply
This failure is confirmed
automatically when the
level has risen to
NIV2
.
The message will
continue to be shown on
the display, however.
Contact the service
department
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...