Chapter 5: Alarms
Fresenius Medical Care
AquaA
IFU-EN 06A-2021
127
5.9 Alarms and information messages–AquaA2 (option)
An error code marked with „*“ denotes a self-confirming message.
Error code
Visual
indica-
tor
Messages
Cause
Action required
F–01–02–01
Signal:
red
FAILURE Stage 2,
BK I/O bus
– BUS system connection
interrupted
– BUS component defective
Contact the service
department
F–01–02–07
Signal:
red
FAILURE Stage 2,
Communication
(measuring
transducer)
– Communication problem
from/to measuring
transducer B4
– Measuring transducer B4
defective
– Serial RS232 connection
line to the KL6031
measuring transducer
defective
Contact the service
department
F–01–02–08
Signal:
red
FAILURE Stage 2,
measuring
transducer (ADC)
– Reference measurement of
test voltage (1.0 V
DC
) failed
– Measuring transducer B4
defective
– Serial RS232 connection
line to the KL6032
measuring transducer
defective
– Digital output terminal A8
defective
– Line connection between
measuring transducer B4
and analog output terminal
A8 defective
Contact the service
department
F–02–02–01
Signal:
red
FAILURE Stage 2,
Permeate cond.
alarm limit
exceeded
– The dialysis water
conductivity has exceeded
the specified alarm limit.
– Conductivity sensor CDT-Ps
defective
– Measuring transducer B4
defective
Contact the service
department
This failure is
confirmed
automatically when
the value drops
below the alarm
limit. The message
will continue to be
shown on the
display, however.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...