Chapter 4: Operation
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
67
4.8.7 AquaUF EMERGENCY MODE (option)
In the event of a malfunction, the
AquaUF
module can be bypassed.
To reroute the dialysis water supply, carry out the following steps:
Step 1
Switch the
AquaA
to
STANDBY
and wait for
STANDBY–Active
mode.
Step 2
Turn off
AquaA
,
AquaA2
and
AquaHT
and at the main power switch
(
OFF/O
position).
Step 3
Open and remove the clamp on the left pipe elbow.
Step 4
Open both clamps on the center pipe elbow.
Step 5
Remove the center pipe elbow.
Step 6
Swivel the left pipe elbow on the front to the right and close it again
using the clamp.
Warning
Risk of contamination
After servicing the dialysis water circuit, the device must be disinfected.
Note
Tighten the clamp connections using 2.5 Nm of torque.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...