Chapter 12: Specifications
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
169
12.8 Transport/storage
For further information (see chapter 10 on page 153)
12.9 External connection options
Other, additional equipment connected to this device must comply with
the applicable IEC or ISO standards (e.g., IEC 60950-1 for information
technology equipment).
Furthermore, all system configurations shall comply with the
requirements for medical systems (see Annex I to ANSI/AAMI
ES60601-1).
Connecting the device to an IT network that contains components not
installed and validated by the manufacturer can introduce unknown
risks for patients, operators or third parties. These risks must be
identified, analyzed, evaluated and monitored by the responsible
organization. For assistance, consult IEC 80001-1 and Annexes H5 and
H6 to ANSI/AAMI ES60601-1.
Any modification to an IT network that has been installed and validated
by the device manufacturer can introduce new risks and therefore
require a repeat analysis. Especially problematic activities include:
– Modifications to the IT network configuration
– Connection of additional components and devices to the IT network
– Removal of components and devices from the IT network
– Updates or upgrades of components and devices in the IT network
Note that local laws take priority over the above-mentioned
requirements. If in doubt, contact the local service department.
Device connections
Ethernet (TCP/IP)
Interface for the exchange of data. Electrically isolated by transformer.
Port:
RJ45
Only systems complying with the regulations or IEC 60950-1 may be
connected to the LAN ports.
Feed flow
FL–F
4.0 to 160.0
L/min
±10 %
Diversion
FL–C
4.0 to 160.0
L/min
±10 %
*MV = measured value, actual value
Measured value
Sensor
Measuring
range
Unit
Accuracy
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...