Chapter 15: Appendix
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
235
Illustration
Description
Fig. 1 - disinfecting the sampling valve:
Disinfect the sampling valve using an alcohol-based skin
disinfectant (without lipid replenisher).
Remove any contamination with a swab.
Then repeat the disinfection procedure. (
Fig. 1
)
Caution:
Observe the acting time of the disinfectant!
Fig. 3 - connecting and locking the adapter:
Place the adapter of the sampling bag onto the sampling valve.
(
Fig. 2
)
Then lock the adapter. (
Fig. 3
)
The multiway valve on the sampling set must be set so as to
ensure that no fluid can flow.
Fig. 4 - opening the sampling valve
Turn the sampling valve counter-clockwise to open it. (
Fig. 4
)
Fig. 1
Fig. 2+3
Fig. 4
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...