Chapter 12: Specifications
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
179
An overvoltage protection device must be installed to prevent damage
to the fuse in the power distribution box of the
AquaA
system. This can
occur when a surge impulse is caused by an atmospheric source, such
as a thunderstorm, or by an unstable power supply.
When using fuses, replacement is recommended every 24 months as
part of the periodic maintenance (MA) procedures.
The use of 3-pole circuit breakers is recommended.
Fuses
Information on electromagnetic compatibility (IEC 60601–1–2)
System type
900
to
3600
Line voltage
208 V, 60 Hz
Power supply
80 A
Tripping characteristic C, D, K or
comparable
Power consumption
22 kVA
Power line
impedance
< (0.15 + j0.15) ohm
Warning
Risk of injury caused by electric shock
Without a protective ground connection, there is a risk of electric shock.
Always connect the device to a power supply network with a
protective earth conductor.
Part number
Fuse
Glass fuse 5 x 20, 5 A T
ATOF
fuse DIN 72581-3C 2A
Tip
The technical specifications are identical with the specifications for the
AquaA
.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...