Chapter 14: Options
200
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
General heat disinfection phases
Filling tank
The filling and reheating of the
AquaHT
tank is independent of the
selected type of heat disinfection.
Heating tank
Ring main heat disinfection
Example temperature graph during ring main heat disinfection:
The temperature graph shows the temperature over time during ring
main heat disinfection:
If the tank level is below the required fill
volume, the tank is replenished using dialysis
water from the
AquaA
. In this step, the
AquaA
is switched to a fill mode and supplies
only the
AquaHT
with dialysis water.
Before each heat disinfection program, the
contents of the
AquaHT
tank are reheated, if
required. When the programmed tank
temperature has been achieved, the actual
heat disinfection program will be started.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...