5 System Operations
24
Fig 5-17
Remote Control Page
In the remote control page in Fig 5-17, you can select the command of “SO2 calibration” in
the list of “Remote Control Command”, as shown in Fig 5-18.
Fig 5-18
Pre-Calibration Page
Click the “Prepare” button to open the calibration electro-valve. Then adjust the flow rate
meter for the standard gas at bottom left of the cabinet front panel to control the flow rate at
(1.5~2) L/min and click the “OK” button to start remotely calibration.
5.3.4.3
Auto Calibration on CEMS-2000 Monitoring Software
You can configure the auto calibration parameters on CEMS-2000 monitoring software and
the system will periodically perform the calibration according to the settings. You can click