7 Operations of CEMS2000 Monitoring Software
76
FOCUSED PHOTONICS INC
7.2.2.5
Configuration of Linear Coefficient
This function is used for configuring the linear coefficient of the measuring items, as shown
in Fig 7-29.
Fig 7-29
Configuration of Linear Coefficient
Please follow the steps below to configure the linear coefficient:
Step 1
Select a measuring variable in Fig 7-29 and input the measured value and
correction value. Click the
button to move the variable to the right
column of “To be fitted”. You should at least input two groups of data and then
click the
button. The system will calculate the data and display it in the
left “Fitting Coefficient”, which contains “primary coefficient” and “zero
coefficient”.
Step 2
Click the “Save” button to save the current settings and click the “Close” button to
return to the main window.
7.2.2.6
Configuration of Fixed Value
This function is used for setting a fixed value for a measuring variable that displays this value
in the main window and stores it in the database, as shown in Fig 7-30.