9 Alert Information and FAQ
109
FOCUSED PHOTONICS INC
9
Alert Information and FAQ
9.1
Alert Information
Alert Code
Content
Level
Measures (Mainly for Protection)
01
Mismatch of sensor type and
pixel
Failure
The spectrometer already runs at
default configuration. You need to
re-configure the light source byte
02
Over limit of optical channel
1
Failure
The spectrometer already runs at
default configuration. You need to
re-configure the optical channel 1
03
Over limit of optical channel
2
Failure
The spectrometer already runs at
default configuration. You need to
re-configure the optical channel 2
04 EEPROM
error
Failure
Restart the system. If the problem
still cannot be solved, you should
change the EEPROM
05
Step motor error
Failure
Restart the instrument. If the
problem still cannot be solved,
please check the relevant circuit of
the motor, the motor and the optical
coupling.
06
Error in EEPROM’s vector
index
Failure
The vector table should be
configured
11
Error in display board
E2PROM
Failure
12
Error in display board RTC
Failure
13
Gas room temperature over
limit
Failure
14
Gas pipe jammed
Failure
15 Temperature
control
error
Failure
16 Pressure
error
Failure
17
SO2 density alert
Warning
18
NO density alert
Warning
19
O2 density alert
Warning