6 Operations of OMA-2000 Spectroscopic Gas Analyzer
52
FOCUSED PHOTONICS INC
Fig 6-35
Zero Adjustment / Calibration Setup
The meanings of items on this page are:
z
Calibration density: Set the density of the standard gas within the range of (0~50000)
ppm
;
z
Manual zero adjustment: Perform manual zero adjustment operations.
Please follow the steps below for manual zero adjustment:
Step 16
Press the “F2” button to enter the manual zero adjustment page, as shown in Fig
6-36.
Fig 6-36
Zero Adjustment
Step 17
Press the “Enter” button on the page of zero adjustment to start adjustment, as
shown in Fig 6-37.