5100ex Calibration Manual
AutoCalibration of Direct Liquid
45
Troubleshooting Direct AutoCal
Error Message
Condition
Possible Solutions
Time dispensed is 0. AutoCal Canceled.
The system was not
run.
Press the green,
standby pushbutton
to dispense liquid.
Pulses measured is 0. AutoCal Canceled.
No pulses were
detected.
Verify the connection
to the direct
feedback sensor.
Value below min. Setting to min.
An inaccurate
number of pulses
were detected.
An inaccurate weight
was entered.
Verify the integrity of
the direct feedback
sensor.
Return to the
AutoCal weight entry
screen and re-enter
the value.