5100ex Calibration Manual
AutoCalibration of Granular Materials
35
Troubleshooting Granular AutoCal
Error Message
Condition
Possible Solutions
Time dispensed is 0. AutoCal Canceled.
The system was not
run.
Press the green,
standby pushbutton
to dispense material.
Pulses measured is 0. AutoCal Canceled.
No pulses were
detected.
Verify the connection
to the auger
feedback sensor.
Value below min. Setting to min.
An inaccurate
number of pulses
were detected.
An inaccurate weight
was entered.
Verify the integrity of
the auger feedback
sensor.
Return to the
AutoCal weight entry
screen and re-enter
the value.