1586A
Users Manual
7-8
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 306
Not waiting for
reference
Cause:
CALibrate:NEXT command with a reference value
was received, but the reference value is not allowed for
current step.
Solution:
Check the parameter value and try again.
Error 307
Continue command
ignored
Cause:
CALibrate:NEXT command was received, but it is not
expected for current calibrate step.
Solution:
Check the parameter value and try again.
Error 318
Reading does not
match reference
Cause:
The measured value out of range, it might be caused
by wrong input signal or hardware defect.
Solution:
Make sure that the input signal matches the
requirement. If the input signal is correct, cycle the power. If
the error occurs again, contact Fluke (see Chapter 1).
Error 320
Timeout waiting for
reading
Cause:
Timeout while waiting readings from inguard.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 322
Password missing
Cause:
CALibrate:ENABle command was received to unlock
the calibration but ADMIN password is not provided.
Solution:
Make sure that ADMIN password was sent with the
CALibrate:ENABle command.
Error 323
Wrong password
Cause:
CALibrate:ENABle command was received to unlock
the calibration but the ADMIN password is wrong.
Solution:
Make sure that correct ADMIN password was sent
with the CALibrate:ENABle command.
Error 324
Calibration locked
Cause:
The calibration is locked when a CALibrate command
was received.
Solution:
Use the CALibrate:ENABle command to unlock the
calibration first.
Error 325
Cannot get stable
readings
Solution 1:
Check calibrator and make sure that the correct
signal is being outputted from the calibrator.
Solution 2:
Cycle the power and run calibration again. If the
error occurs again, the relay card needs serviced, contact
Fluke (see Chapter 1).
Error 398
Anomalous error %d
Cause:
Unknown error occurred during the calibration
procedure.
Solution:
Restart the calibration procedure. If the error
occurs again, contact Fluke (see Chapter 1).
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...