Error Messages and Troubleshooting
Introduction
7
7-3
Introduction
This chapter supplies information on error messages and how to troubleshoot the product.
Error Messages
Error messages help the user diagnose problems with the Product. Error messages contain
a message along with an error code. If an error message shows on the display, use the
information in Table 7-1 to resolve the problem.
Table 7-1. Error Messages
Error Code
Error Message
Causes and Solutions
Error -350
Queue overflow
Cause:
The error queue is full because more than 16 errors
have occurred.
Solution:
The error queue is cleared by *CLS (clear status)
command or when power is cycled. The errors are also
cleared when the queue is read.
Error -224
Illegal parameter
value
Cause:
A discrete parameter was received that was not a
valid choice for this command.
Solution:
Invalid parameter value used. Check the
parameter values and try again.
Error -222
Data out of range
Cause:
A numeric parameter value is outside the valid range
for this command.
Solution:
Check the parameter value and try again.
Error -213
INIT ignored
Cause:
An INITiate command was received but could not be
executed because a scan was already in progress.
Solution:
Send an ABORt command to stop a scan in
progress.
Error -211
Trigger ignored
Cause:
More than one trigger was received while the
instrument was scanning.
Solution:
Make sure the proper trigger type is selected. If the
problem persists, try to slow down the rate that the trigger
occurs.
Error -171
Invalid expression
Cause:
The expression data element was invalid. This could
be because of an unmatched parentheses or an illegal
character.
Solution:
Check the parameter value and try again.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...